Not Another Name, Not Another MRI Accident

B. Rahimpour, Smith Chason College

School of Medical Imaging

Lessons from 1980 to 2001 and 2025

The first clinical MRI scan in 1980 was a watershed event, often compared in importance to the discovery of X-rays in 1895, It demonstrated conclusively that MRI was not just a laboratory experiment but a powerful medical imaging tool capable of detecting tumors and other internal abnormalities without surgery or ionizing radiation. Following the Aberdeen achievement, MRI technology advanced rapidly.

MRI (Magnetic Resonance Imaging) is an invaluable diagnostic tool, yet if safety protocols are not strictly followed, it can turn deadly. Two tragic incidents, 24 years apart, underscore that MRI accidents still occur despite advances in technology and decades of experience. On July 31, 2001, 6-year-old Michael Colombini was killed during an MRI scan when a portable oxygen tank was brought into the scan room and became a deadly projectile. On July 17, 2025, 61-year-old Keith McAllister died after his heavy metal necklace was pulled into an MRI at an outpatient center on Long Island. These cases—one involving a high-field superconducting magnet in 2001 and the other an Open MRI facility in 2025—show that no matter the magnet type, location, or the patient’s age, catastrophic accidents can happen if safety procedures are ignored. The only factor that can prevent “not another name, not another MRI accident” is rigorous knowledge and adherence to MRI safety policies at all times.

Michael Colombini’s 2001 MRI Accident

Westchester Medical Center

On July 26, 2001, 6-year-old Michael Colombini was undergoing a routine follow-up MRI after brain tumor surgery at Westchester Medical Center in New York. Partway through the scan, the child’s oxygen supply faltered. In the scramble to provide oxygen, a staff member brought a standard steel oxygen cylinder into the MRI suite, unaware (or forgetful) of the magnet’s powerful pull. In an instant, the 10-ton MRI magnet (approximately 30,000 times Earth’s magnetic field) yanked the heavy oxygen tank across the room, a phenomenon known as the “missile effect.” The airborne cylinder struck Michael’s head, causing a fatal skull fracture. Despite surgical intervention, Michael died from his injuries two days later.

This was the first known MRI-related death due to a projectile, and it sent shockwaves through the radiology community. Investigations revealed multiple safety failures leading up to the accident. The MRI suite’s built-in oxygen system had malfunctioned, prompting staff to seek alternative oxygen. Both technologists on duty left the control room to address the oxygen supply issue, leaving an anesthesiologist alone with the sedated child. In the urgency, no one stopped the ferromagnetic tank from being carried into the magnet room—a critical lapse in screening and training. The tragedy highlighted factors such as poor planning, improper emergency procedures, inadequate staff training, and failure to segregate MRI-unsafe equipment. As MRI safety expert Tobias Gilk later summarized, “for all our talk about MRI safety,” the community had failed to extract the hard lessons from this tragedy at the time.

Michael Colombini’s death became a somber turning point for MRI safety awareness. The American College of Radiology (ACR) had already been developing safety guidelines, but this incident galvanized efforts to prevent such an accident from ever happening again. The following year, the ACR issued its first “White Paper on MR Safety,” and over the next two decades the field saw major initiatives to standardize MRI safety protocols. Westchester Medical Center accepted full responsibility for the accident, and the Colombini family later reached a legal settlement of $2.9 million in 2010. More importantly, Michael’s legacy prompted hospitals worldwide to re-examine their MRI safety practices—from ensuring only MRI-compatible oxygen tanks and equipment are used, to instituting emergency drills and redundancies for situations like oxygen failures. Yet, as we’ll see, even “never again” proved not to be never, because the lessons of 2001 must be continually enforced to truly prevent a repeat.

Keith McAllister’s 2025 MRI Accident

Nassau Open MRI

In July 2025—44 years into the MRI era—a fatal accident eerily echoed the Colombini case, despite the very different setting. Keith McAllister, 61, accompanied his wife to an outpatient imaging center (Nassau Open MRI in Westbury, NY) for her knee MRI exam. The facility used an “open” MRI scanner, which typically has a lower magnetic field strength than traditional closed-bore MRI units. Tragically, field strength made little difference; the magnet was still powerful enough to turn a 20-pound metal chain around Keith’s neck into a lethal projectile.

According to reports, after the scan was completed the MRI technologist allowed Keith into the scan room (Zone 4) to help his wife off the table, despite him wearing a large steel necklace. The moment Keith stepped near the magnet, the MRI’s force “snatched” the chain and pulled him in violently. He was slammed against the machine, the chain reportedly tightening around his neck and torso. Keith’s wife Adrienne, who witnessed the horror, recalled that “in an instant, the machine switched him around, pulled him in, and he hit the MRI,” at which point “he went limp” in her arms. Panicked, she and the technologist struggled to free him, and she pleaded for someone to “turn this damn thing off!”

Emergency responders arrived to find Keith in cardiac arrest. He suffered multiple heart attacks on site and en route to the hospital. Sadly, he succumbed to his injuries the next day, July 17, 2025. The investigation revealed that this was not a case of an intruder wandering in unescorted—the staff had actually permitted him to enter with the metal chain on, a chain they’d noticed on prior visits and even joked about due to its size. In other words, basic MRI screening protocols were flagrantly violated. Patients and accompanying persons are always supposed to remove all metal objects and be safety-screened before entering the magnet room, yet that did not happen here.

Keith McAllister’s death made headlines as a “freak MRI accident,” but MRI professionals noted it was disturbingly familiar. News of the case cited the Colombini incident of 2001 as an earlier parallel in New York where a projectile caused a death. The exact same known projectile effects and the very main cause. In the wake of the McAllister accident, we could find footprints of lack of enough knowledge and lack of adherence to firmly and fully expressed guidelines for Safe MR Practice published by ACR and MRI experts in the MRI field.

Why Do MRI Accidents Still Happen?

It may seem shocking that in 2025, with MRI technology in use for over 40 years, such grave safety lapses can still occur. The unfortunate truth is MRI accidents have persisted at a high but steady rate and recently have increased in frequency, usually for the same reasons: human error, complacency, or ignorance of safety protocols. Dr. Emanuel Kanal, a renowned MRI safety expert, notes that while MRI is generally very safe, “when someone is hurt during MRI, the injury was almost always avoidable.” In other words, virtually every MRI accident is preventable with proper precautions. Dr. Kanal emphasizes that incidents typically result from a “lack of knowledge about MRI safety protocols and guidelines” among those involved.

Despite the MRI community’s efforts since 2001 to promote safety, gaps in training and awareness remain. In many countries and facilities, there are no legally mandated MRI safety standards—best practices exist (via organizations such as ACR, Joint Commission, etc.), but compliance often relies on voluntary adherence. This can lead to uneven enforcement. For example, a busy outpatient center might grow complacent if they’ve never seen an accident, leading staff to shortcut screening procedures. In the 2025 case, familiarity and the informal atmosphere (an outpatient MRI center where the patient’s husband was known to staff) likely led to a false sense of security—a belief that “it’s just an open MRI, what could happen?” This culture of complacency can be fatal. The MRI magnet doesn’t distinguish between a children’s hospital or an outpatient clinic, nor between a patient or a visitor: ferromagnetic metal will be violently pulled in regardless.

As in any well-structured academic MRI education it’s emphasized and highlighted: “An MRI scanner is an extremely strong Magnetic System with a Magnet Always On.” Awareness of this fact is crucial; any lapse—even one moment of letting an uncleared person or object into the room—can have dire consequences.

It’s also worth noting other MRI accidents worldwide echo the same patterns. In 2018, a 32-year-old man in Mumbai, India was killed when he carried an oxygen cylinder into an MRI room—again, the tank was pulled from his grasp and caused fatal injuries. In 2021, a hospital patient in South Korea died under similar circumstances when an oxygen cylinder left near the MRI became a projectile. Each time, investigations cite a breakdown in screening and safety procedures. In essence, the MRI industry should emphasize proper MRI education, employment of qualified MRI Technologists as Level II MR Personnel, and mandating MRI Safety requalification as an annual requirement for all MRI Technologists.

Strengthening MRI Safety: Protocols and Best Practices in 2025

An MRI scanner in a clinical setting as any MRI Magnet, regardless of field strength, can pose deadly risks if safety protocols are not followed. As MR Physics states, “The Magnet is Always On” so all equipment in and around MRI department must be labeled as MRI Safe, Unsafe and Conditional, and carefully checked screened before entering the scan room. Preventing MRI accidents requires a culture of safety and strict adherence to well-established protocols. The good news is that the MRI community through organizations including the ACR, Joint Commission, and the American Board of MR Safety has developed clear guidelines that, if followed diligently, make MRI suites extremely safe. Here are key MRI safety measures and lessons, drawn from these guidelines and past accidents:

  • Four-Zone Controlled Access: MRI facilities are structured into safety zones (I through IV) to control access as one approaches the scanner. Zone 4 is the MRI room itself often called the “Magnet Room” or “Projectile Zone” where the Magnetic Feld is strongest and poses the greatest hazard, Zone 3 (the area just outside the Magnet Room, such as the console area) is strictly restricted to screened individuals access only. To prevent incidents like a ferromagnetic object entering Zone 4, no one unscreened should ever get near Zone 4. In practice, this means locking the MRI room when not in use, and actively controlling who passes through Zone 3. If “someone walks into Zone 3 with an unsafe object and the door to Zone 4 is open,” it sets the stage for disaster. Facilities must enforce that only authorized, trained MRI personnel or screened patients/visitors are allowed in Zones 3 and 4, a rule that was unfortunately broken in the McAllister case.
  • Rigorous Screening of Everyone Entering Zone 4: Every individual who enters the MRI scanner room whether a patient, a family member, staff, maintenance worker or even Radiologist or a patient accompanying Physician must undergo thorough full screening including clearance metal or medical implants. The ACR recommends a two-step screening process: a detailed written questionnaire followed by a verbal interview by trained MRI personnel (Level 2 MRI safety-trained staff) to confirm and clarify any risk factors. Patients typically fill out screening forms covering implants, devices, or metal in their body, and must remove all metallic items (jewelry, piercings, etc.) before imaging. Accompanying family or caregivers should be held to the exact same screening standard as patients before they enter Zone 4. (In fact, ACR guidance states that any non-patient entering Zone III/IV should be screened “using the same criteria as anyone else” and limited to one necessary companion) In both the 2001 and 2025 accidents, this protocol was breached: the oxygen tank in 2001 and Keith’s necklace in 2025 would never have been in the room if proper screening and metal removal had occurred. These tragedies starkly reinforce that skipping the screening process even once is not an option. Modern best practice is also to perform a final “Pause and Check” before crossing the threshold into Zone 4: a last-minute verification that the patient (and anyone with patient) has removed all metal, and nothing has changed since initial screening.
  • “The Magnet is Always On” Signage and Vigilance: MRI magnets, especially superconducting magnets, are always at full field (unless ramped down for service). There is no “off switch” for the static magnetic field in normal operation. This fact must be mastered by all staff and clearly communicated via signage. Many MRI suites post warning signs at the door such as “DANGER: HE MAGNET IS ALWAYS ON” This reminds everyone that even if no scan is running, the magnet will still forcefully attract ferromagnetic objects. In an MRI accident narrative, one often hears someone shout “Turn off the machine!” as in Adrienne McAllister’s cry. However, quenching (emergency shutdown) of the magnet is not instantaneous and doing so carries its own risks and it’s not like flipping a power switch. Prevention by keeping hazards out of the room is the only real protection. Thus, constant vigilance is required: for example, housekeeping or security personnel must also be trained not to inadvertently bring in mop buckets, tools, or firearms into Zone 4. (One reported incident involved a police officer’s gun being sucked onto an MRI magnet during a hospital raid, illustrating that anyone near an MRI needs safety screening. Simply put, all staff and visitors must treat the MRI room as a high-risk area at all times. The culture should be that no one enters Zone 4 without explicit permission from MRI technologists, who act as gatekeepers of Zone 4.
  • MRI-Compatible Equipment Only: Hospitals and imaging centers have learned to segregate MRI Safe vs. Unsafe equipment. Items like oxygen cylinders, wheelchairs, stretchers, IV poles, and monitoring devices must be MRI Safe (non-ferromagnetic) if they will enter Zone 4. A critical error in 2001 was that the facility did not have an MRI Safe oxygen delivery backup readily available, nor were ferrous oxygen tanks clearly marked or secured away from MRI Zones. Today, best practice is to keep a supply of MRI- Safe oxygen tanks (typically aluminum cylinders) or piped-in oxygen in the MRI room and to remove all conventional Stainless Steel tanks from Zones 3 and 4 entirely. Likewise, tools used by anesthesiologists or engineers in MRI should be non-magnetic. Had these precautions been standard in 2001, Michael’s accident might have been averted by simply using a non-ferromagnetic O₂ tank. The same principle applies to personal items: lockers should be provided, and patients/visitors must stow everything (phones, keys, coins, jewelry, etc.) before entering. In Keith McAllister’s case, his 20-lb metal chain was an obvious hazard that should never have been allowed near the magnet. Ensuring that nothing MRI Unsafe ever crosses into Zone 4 is one of the simplest and most effective safety measures, truly a zero-tolerance policy for ferromagnetic.
  • FerroMagnetic Detection systems (FMDs): In addition to manual screening, many facilities now employ ferromagnetic detectors at the MRI room entrance as an extra layer of defense. These detectors act somewhat like metal detectors but are tuned to ferromagnetic materials. The ACR regards them as a supplement not a replacement to traditional screening. A conscientious facility might have patients and staff walk through a portal detector or be wanded by a handheld detector before entering Zone 4. In fact, after the 2018 Mumbai MRI death, it was reported that the metal detector at that site was not functioning a factor that potentially contributed to the accident. While an FMD is not foolproof and can sometimes give false alarms, it can catch a forgotten ferrous object (like a hairpin or belt buckle) that manual screening missed. The cost of these systems is relatively low, especially compared to the human cost of an accident. Many safety experts strongly recommend using FMDs as a standard practice, given that they add an additional safety net against human error.
  • Training and Safety Drills: Perhaps most importantly, MRI safety training must be ongoing for all personnel. This includes not just MRI Technologists, but Radiologists, nurses, anesthesiologists, maintenance staff, and even reception or transport staff who may interact with MRI patients. The goal is to foster a “Safety-first” mindset in all MRI operations. Regular safety drills and refreshers can help. For example, staff should practice what to do in emergencies (quench procedures, response to a fire or code in MRI, etc.) and how to properly screen and escort patients. The depth of knowledge of each individual matters, it’s the cumulative effect of everyone doing their part that prevents accidents. ACR guidelines define levels of MRI personnel training (Level 1 for basic safety knowledge, Level 2 for advanced knowledge) and recommend at least one  Level 2 person (often the MRI technologist) be present whenever scanning is occurring. The importance of empowerment should be stressed too: any staff member should feel enabled to speak up if they notice a potential safety issue (e.g., “Stop, I’m not sure that object is MRI Safe” or “Wait, has this person been fully screened?”). In the end, human vigilance is the strongest defense. As Dr. Kanal has stated many times MRI injuries are “almost always avoidable” it just takes everyone being informed and attentive to avoid them.
  • Learning from Incidents (“Near Miss” reporting): The MRI community encourages a culture of transparency and learning from mistakes. When incidents or “near-misses”
    occur (even if no injury happened), they should be reported and analyzed, not swept under the rug. After 2001, many hospitals implemented incident reporting for MRI hazards for instance, if a paperclip flies into the magnet but harms no one, it’s still taken seriously and used as a teaching moment. Such reporting can identify weak points in the process (Was someone not screened? Did someone bypass a policy? etc.) so that processes can be tightened. Over time, this leads to what aviation calls a “culture of safety,” where everyone is mindful of potential dangers at all times. The establishment of MRI Safety Week (an annual event around the anniversary of Michael Colombini’s accident) is one example of the community’s commitment to keep safety lessons fresh. In July 2025, the International Society for Magnetic Resonance in Medicine (ISMRM) hosted educational workshops during MRI Safety Week, aiming to share best practices and latest knowledge with MRI professionals worldwide, and it’s saddening starting 2025 MRI Society has two reasons to hold MR Safety week in month of July.

In conclusion, the haunting parallels between the 2001 and 2025 MRI accidents teach us that technology alone doesn’t guarantee safety people with their proper education and abiding to
protocols do. It does not matter if the magnet is a 1.5 Tesla superconducting unit or a 0.2 Tesla open scanner; it does not matter if the victim is a patient or a patient’s companion. What
matters is whether stringent safety procedures are in place and followed by every individual who sets foot in the MRI environment. With MRI use now spanning over four decades, our
collective knowledge on how to prevent accidents is extensive. The challenge is ensuring that this knowledge translates into action every single day in every MRI facility. Each name added to
the list of MRI accident victims, whether it’s Michael Colombini, Keith McAllister, or others around the world, is a tragic reminder that complacency or ignorance can be deadly. By contrast, facilities that foster a robust safety culture demonstrate that MRI exams can be conducted for years without a single injury.

Let Michael’s and Keith’s stories sharpen our resolve to enforce MRI Safety with utmost diligence. From the moment a patient is referred to MRI, through screening, to the scan and
beyond, safety must remain the foremost priority. As one MRI safety sign boldly states: “THE MAGNET IS ALWAYS ON” meaning our caution and vigilance must always be “on” as well. If the
radiology community internalizes this lesson, then “not another name, not another MRI accident” can move from a hope to a reality. Each of us Radiologists, Technologists, Nurses, and
Administrators has a role to play in making sure these preventable tragedies never happen again. No scan is worth a life, and with proper precautions, no one should ever be injured by an MRI. The ultimate tribute to those we’ve lost is to redouble our commitment to MRI safety so that history does not repeat itself.

 